Turbidity Sensor (Middle Range) TSC-D Series
High cost performance model with separate transmitter and detector.
Products Name |
---|
TSC-10-D,[Parts No. A10005337] |
Features
![]() |
90 degree scattered light method Turbidity Sensor TSC-10S(E) uses a method of 90 degree scattered light as the measuring principle of the world common standard. |
![]() |
Built-in wiper cleaning system The built-in wiper cleaning system certainly keeps lenses clean. |

![]() |
Sapphire glass of the optical windows The optical windows are made of hard-to-scratch sapphire glass. It enables to scrub the window surface to maintain the Turbidity Sensor. |

![]() |
Compact design The compact design makes it possible to install easily. |

Application
Measurement of the effluent
Measuring the turbidity of the effluent from the waste water treatment plant makes it possible to monitor the treatment condition.

Measurement of the intake water
Measuring of the intake water for drinking water manufacturing process makes it possible to prevent any troubles that may be caused by sudden turbidity change.

Installation
Installation example

Extension of sensor cable
The standard length of the sensor cable is 9 m. To extend the cable, refer to the following.
To extend the cable, connection using pull box is recommended.

Transmitter
Display item
Set the unit of turbidity to show on the SC-U1 display.


Options

Mounting attachment: TA-3
[Parts No. 05075]
By mounting the turbidity sensor and SS sensor to the pipe, it prevents hitting the wall or facility equipment and rotating it. Please use it when the water current is fast.
Applicable pipe: Stainless steel pipe for piping 25 A (OD 34 mm)
Material: S30400
*Please prepare the pipe for installation by customer

Maintenance kit: TSC-MK
[Parts No. 05076]
It is a replacement blade of the wiper cleaning device. Please exchange once a year as a guide.
Component
●Wiper blade × 2
●Cleaning brush × 1
●Detection window Cleaning paper × 2
●Manual × 1
Specifications
Dimensions
Model | TSC-10-D |
---|---|
Measurement range | 0.00 – 1000.0 NTU (formazin standard solution) |
Display resolution | 0.00 – 19.99 NTU: 0.01 NTU 20.0 – 1000.0 NTU: 0.1 NTU |
Power supply voltage | 24 VDC ± 10% |
Current consumption | Standard: 30 mA max. Cleaning: 230 mA max. |
Communications (function) | MODBUS (RTU/ASCII) |
Cleaning system | Wiper cleaning system |
Operating temperature | 0 to 40°C (unfrozen) |
Major material | SUS316L, Sapphire glass, Fluorocarbon rubber, EPDM, PVC (cable) |
Dimensions | Approx. Φ48 × 133 mm |
Weight | Approx. 1.1 kg |
Degree of protection | IP68, maximum depth of 2 meters (underwater type) |
Sensor cable length | 9 m |