Turbidity Sensor (Middle Range) TSC Series
High cost performance model with separate transmitter and detector.
Features
- 90 degree scattered light method
Turbidity Sensor TSC-10(E) uses a method of 90 degree scattered light as the measuring principle of the world common standard. - Built-in wiper cleaning system
The built-in wiper cleaning system certainly keeps lenses clean. - Sapphire glass of the optical windows
The optical windows are made of hard-to-scratch sapphire glass. It enables to scrub the window surface to maintain the Turbidity Sensor. - Compact design
The compact design makes it possible to install easily. - Approval Certificate
Turbidity Sensor TSC-10(E) got the Pattern Approval Certificate of the measuring instruments of the People's Republic of China.
Application
Measurement of the effluent
Measuring the turbidity of the effluent from the waste water treatment plant makes it possible to monitor the treatment condition.

Measurement of the intake water
Measuring of the intake water for drinking water manufacturing process makes it possible to prevent any troubles that may be caused by sudden turbidity change.

Options

Mounting attachment: TA-3
[Parts No. 05075]
By mounting the turbidity sensor and SS sensor to the pipe, it prevents hitting the wall or facility equipment and rotating it. Please use it when the water current is fast.
Applicable pipe: Stainless steel pipe for piping 25 A (OD 34 mm)
Material: S30400
*Please prepare the pipe for installation by customer

Maintenance kit: TSC-MK
[Parts No. 05076]
It is a replacement blade of the wiper cleaning device. Please exchange once a year as a guide.
Component
●Wiper blade × 2
●Cleaning brush × 1
●Detection window Cleaning paper × 2
●Manual × 1

Stanchion pole: PS-1
[Parts No. 05031]
If there is no place (such as a wall) where the transmitter is installed, it can be mounted on the pole together with the hood kit.
Material: S30400

Hood kit: TP-FK2
[Parts No. 05020]
It is a sunshade cover that protects the transducer from direct sunlight.
Material: S31603
Specifications
Dimensions
Model | TSC-10(E) | |
---|---|---|
Measuring range | 0.00–500.0 (NTU/FNU) | |
Power supply voltage | 100–240 VAC ±10% 50/60 Hz | |
Power consumption | Normal: 15 VA or less, During cleaning: 22 VA max. | |
Display resolution | 0.01 (0.00-19.99), 0.1 (20.0-500.0) NTU/FNU | |
Output | Signal Output | analog 4-20 mA, resistance load of 300 Ω max. |
Self-checking Relay Output | non-voltage C-contact capacity 240 VAC, 1 A resistance load | |
Alarm Relay Output | non-voltage C-contact capacity 240 VAC, 1 A resistance load | |
Alarm timer | 1 to 120 min. (adjustable) | |
Calibration | Distilled water | |
Cleaning system | Automatic wiper cleaning system | |
Time interval for cleaning | 10 to 240 min. (selectable) | |
Measuring water temperature | 0 to +40°C (unfrozen) | |
Ambient temperature | Transmitter:-20 to +50°C, humidity 95%RH or less (Avoid direct sunlight) | |
Operating altitude | Altitude up to 2000 m | |
Main material | Detector: SUS316L, sapphire glass, fluorocarbon rubber, EPDM, Polyolefin (cable) | |
Transmitter: Polycarbonate | ||
Dimensions | Detector: Approx. Φ48 x 133 mm (Approx. Φ1.89 x 5.24 in) | |
Transmitter: Approx. 240 (W) x 162 (H) x 75 (D) mm [Approx. 9.45 (W) x 6.38 (H) x 2.95 (D) in] | ||
Weight | Detector: Approx. 1.0 kg (Approx. 2.20 lb) | |
Transmitter: Approx. 1.6 kg (Approx .3.53 lb) | ||
Degree of protection | Detector: IP68, maximum depth of 2 m (underwater type) | |
Transmitter: IP65 (jetproof type) Pollution degree 2 | ||
Detector cable length | 9 m |
Detector

Transmitter

Contact phone number
- Headquarters
- +81-77-579-8680
Contact form
Support information