PT-S80/U80 back
Outside dimensions options coution

specification
Model
PT-S80 PT-U80
Measuring range -30 to 600°C
Field of view 30mm/1000mm [D:S=33:1]
Optics Si lens
Sensing element Thermopile
Wavelength 8 to 14 μm
Response time 0.5 sec. / 90 %
Accuracy (ε0.95) −30.0 to 0°C:3°C,0.1°C to 200°C:2°C,201°C to 600°C:1%
Repeatability 1°Cof reading value
Display resolution -30.0 to 199.9ºC: 0.1ºC, 200 to 600ºC: 1 %
Sighting method Coaxial Laser marker (Class 2)
HOLD time 15 seconds
Continuous
measurement mode
N / A ON/OFF Selectable
USB output N / A Available
Data storing
Single memory 35-point memory
(max 150-point)
Back light EL back light
High / Low alarm Alarm LED, Buzzer, (ON/OFF Selectable)
Emissivity (E)
adjustment
0.95 / 0.85 / 0.70 Selectable ε(0.3 to 1.20 / 0.01 step)
Display function NOR/MAX/MIN
Power supply AA alkaline battery (2 pieces)
Battery life 15 hours (with max load)
Ambient
temperature
0 to 50°C
Ambient humidity 35 to 85 % RH (no condensation)
Storage temperature
/humidity
−10 to 60°C/35 to 85% RH
Material ABS/TEEE
Dimensions H×W×D = 182×56×38mm
Weight Approx. 230 g (including batteries)
Accessories: Size AA alkaline battery (two) for operation check, Instruction manual, Protective case for PT-S80/U80, USB cable (for PT-U80 only), Optional: Black body tape HB-250
Specifications are subject to change for product improvement without prior notice.

up


Outside dimensions
Outside dimensions

up

standard accessory
Protective case



options
HB250
HB-250
The Black Tape (HB-250) is designed for more accurate measurement, especially if the target object has a shiny surface. Apply HB-250 on the surface of the target and measure the area covered by HB-250 with emissivity setting at 0.95.

up


caution
Difficult measurement cases
· When the surface being measured is reflective, such as polished metal.
(Before measuring, apply the optional HB-250 tape or apply a matte finish.)
· When measuring through glass

Safety instructions
· Please read through the instruction manual before use.
· This product is not a clinical thermometer, therefore, cannot be used for medical purposes.
· Keep the thermometer away from water contact, as it is not waterproof.
· Avoid rapid change of the ambient temperature, as this may cause measurement error.
· eep the thermometer away from strong electromagnetic sources.

Laser beam
· This product uses laser classified as class II complying with JIS C 6802.
· For safe use, follow the warning labels attached to this product.
· This is a portable product with the laser maker certificated by JQA (Japan Quality Assurance of Organization).

up